| The following is the general formula for determining necessary heating and cooling requirements of a system. | |||||||||
| Formula: Q=mC(T2-T1)/t | |||||||||
|
where: |
m= |
mass of bath/reactor fluid | |||||||
|
C= |
specific heat of bath/reactor fluid | ||||||||
|
T1= |
start temperature | ||||||||
|
T2= |
end temperature | ||||||||
|
t= |
elapsed time | ||||||||
| Directions: | Please substitute your own values as indicated below. | ||||||||
|
Value |
|||||||||
| 1 | 20 | liters | Enter the volume of the circulator + hoses + reactor jacket volume | ||||||
| 2 | 0 | liters | Enter the volume of the reactor (enter zero if not applicable) | ||||||
| 3 | 1 | kg/liter | Enter the density of the circulator bath fluid | ||||||
| 4 | 0 | kg/liter | Enter the density of the reactor fluid (enter zero if not applicable) | ||||||
| 5 | 1.45 | kJ/(kg°C) | Enter the specific heat capacity of the circulator bath fluid | ||||||
| 6 | 0 | kJ/(kg°C) | Enter the specific heat capacity of the reactor fluid (zero if not applicable) | ||||||
| 7 | 120 | °C | Enter the starting temperature | ||||||
| 8 | -10 | °C | Enter the target temperature | ||||||
| 9 | 14000 | seconds | Enter the desired heat-up/cool-down time | ||||||
| 10 | kW | Enter estimated power losses (this number must be positive) | |||||||
| Bath Liquid Properties | |||||||||
| Value | m1(kg) | ||||||||
| 11 |
20 |
Multiply value 1 times value 3 and enter result here | |||||||
| C1 (kJ/[kg°C]) | |||||||||
| 12 |
1.45 |
Enter value 5 here | Reference Data | ||||||
| Reactor Liquid Properties | Specific Heat | Density | |||||||
| m2 (kg) | kJ/(kg°C) | kg/liter | |||||||
| 13 |
|
Multiply value 2 times value 4 and enter result here | water | 4.18 | 1 | ||||
| ethanol | 2.43 | 0.783 | |||||||
| C2 (kJ/[kg°C]) | methanol | 2.49 | 0.787 | ||||||
| 14 |
|
Enter value 6 here | silicon oil | 1.45 | 1 | ||||
| aluminum | 0.896 | 2.7 | |||||||
| Customer Requirements | copper | 0.385 | 8.9 | ||||||
| Value | t (s) | iron | 0.45 | 7.84 | |||||
| 15 |
14000 |
Enter value 9 here | lead | 0.128 | 11.31 | ||||
| silver | 0.235 | 10.47 | |||||||
| T1 (°C) | tin | 0.217 | 5.73 | ||||||
| 16 |
120 |
Enter value 7 here | ammonia | 4.8 | 0.602 | ||||
| mercury | 0.139 | 13.56 | |||||||
| T2 (°C) | 50/50 EG/H20 | 3.25 | 1.1 | ||||||
| 17 |
-10 |
Enter value 8 here | |||||||
| Required Heating/Cooling Power | Modified Equation: | ||||||||
| Q=[(m1C1)+(m2C2)][(DT)/t] | |||||||||
| Temperature Change | |||||||||
| Value | DT (°C) | ||||||||
| 18 |
-130 |
Subtract value 16 from value 17 and enter result here | |||||||
| 19 | Nominal Heat Load (kW) | Nominal Heat Load is the calculated load based on the customer's requirements. | |||||||
|
-0.269 |
Perform the following calculation: [(v11v12)+(v13v14)][(v18)/v15] | ||||||||
| 20 | Total Heat Load (kW) | Total Heat Load is the Nominal Heat Load plus any estimated losses. | |||||||
|
-0.269 |
If value 19 is positive, add value 10. If value 19 is negative, subtract value 10. | ||||||||
| Heating or Cooling? | |||||||||
| 20 |
Cooling! |
If value 19 is positive, heating is required. If value is negative, cooling is required. | |||||||
| Important Note: This calculation should only be used as a basis for circulator/chiller selection. Many factors that will influence power requirements are too difficult to quantify, and thus are neglected from this calculation. Actual power requirements will almost always be larger than calculated values. | |||||||||
| Additional Notes: | |||||||||
| 1 | Julabo is not responsible for damages/injuries caused by using unapproved bath fluids in circulators/baths/chillers. | ||||||||
| 2 | Julabo is not responsible for damages/injuries resulting from incorrect use of approved bath fluids. | ||||||||
| 3 | Flash points of bath fluids must be above +60°C for transport by plane. | ||||||||
| 4 | Julabo chiller/circulator high temperature warning and safety values should reference fluid flash temperature. | ||||||||
| 5 | Julabo chiller/circulator low temperature warning should reference fluid freezing point or maximum viscosity. | ||||||||
| 6 | Julabo H350 bath oil is intended only for Julabo high temperature sealed systems. | ||||||||